Agilent 6120 Single Quadrupole Walkup instrumentation is a simple and efficient tool to perform complex LC/MS analysis in a typical multi-user environment. The administrator privileges within Agilent MassHunter Walkup Software enable the administrator to create Walkup methods, maintain instrument performance, and manage users. The administrator can add users and assign them different capabilities and access rights. MassHunter Walkup software can be integrated with Agilent OpenLAB services for maintaining confidentiality in data management and results in distribution according to lab protocols. An administrator can prepare the Walkup software as a ready-to-use system to perform various advanced tasks. These tasks can include, for example, compound confirmation, purity checking, and identification of impurities, and can be performed by a non-expert user.

Technical Data

Mass Range: m/z 10 - 1500

Mass Accuracy: ± 0.13 u within the calibrated mass range in scan mode

Mass Axis Stability: The mass drift does not exceed the larger absolute value of ± 0.1 Da or ± 100 ppm of measured mass over a 12 hour period at constant temperature ± 3 degrees centigrade.

Scan Speed: 2500 u/s

SIM Sensitivity: 10 pg reserpine, 100:1 RMS (20:1 peak-to-peak)

Multisignal Acquisition: Ability to cycle through four acquisition modes on a scan-by-scan basis within a single run. Up to four analysis modes can be viewed in the software window and can be linked together to facilitate processing and analysis. Modes of operation: positive/negative polarity switching, variable fragmentor voltage selection, and concurrent SIM/scan.

Demo

 
KRSS Left Strip