ZN2-HP

The F-DGSi High Purity (HP) Nitrogen Generators are engineered to transform standard compressed air in to a safe regulated supply of 99.9995% pure nitrogen, O2 and moisture free and a catalyst to remove hydrocarbons (CH4) < 0.05 ppm from Nitrogen.
  • Advantages
  • Characteristics
  • Specifications
  • Demo
  • Quote

Advantages

Increased laboratory efficiency
A constant, uninterrupted gas supply of guaranteed purity eliminates interruptions of analyses to change cylinders and reduces the amount of instrument re-calibrations required.

Improved economy
Zero Nitrogen gas produced as standard

Improved safety
Zero Nitrogen produced at low pressure and ambient temperature removes the need for high pressure cylinders

Simple installation
Gas generators can be installed in the laboratory, on or under a bench, eliminating the need for long gas lines from cylinders secured elsewhere

Characteristics

  • Fully regenerative PSA technology ; reduced risk of gas contamination and phthalate free
  • Generates Nitrogen on demand from external compressed air
  • Catalyst module for hydrocarbons level (CH4)
    < 0.05 ppm
  • Designed to run 24 hours a day
  • Smart control of the flow of inlet air, it is function of the real flow of nitrogen used by the application
  • Minimum maintenance with an annual filter change
  • Remote PC monitoring via USB/RS232.
  • Ultra fast start-up time and fast purity
  • Compact and modular design

Specifications

Model
ZN2-600-HP.D
N2 outlet
Flow rate (max)
600 cc/min
Outlet pressure (min)
1 bar (14.5 psi)
Outlet pressure (max)
Inlet pressure 2 bar (29 psi)
Nitrogen purity
> 99.9995% (< 5ppm O2 @ 7 Bar(101.5 psi) inlet pressure)
O2 content
< 5 ppm
Hydrocarbon (CH4 free)
< 0.05 ppm

Air inlet
Air inlet quality requirement
Particulate : < 1 micron
Dewpoint : -40°C (-40°F)
Oil < 0.01 mg/m3
Flow rate (max)
15 l/min
Supply pressure (min - max)
7 bar(101.5 psi) 10 bar (145 psi)
Temperature (min - max)
1°C (3°F) -35°C (95°F)

Communication
LCD touch screen
Yes
RS232/RS485/USB
X

General Data
Power supply
100 - 240 V ac / 50 -60 Hz 2 A
Connection type
IEC320-C13
Installation power (max)
240 W (280 V ac)
Fuse rating
4 A 240 V ac (5 x 20 mm)
Net weight (kg/lbs)
< 25 / < 55
Dimensions (W x H x D)
30 x 47 x 43 cm (11" x 18" x 17")

Connections
Outlet/Inlet port
G 1/4" Female

Operating conditions
Temperature
5 - 35°C (41 - 95°F)
Humidity (max, non condensing)
80% at 25°C (77°F)
Noise
< 25 dB(A)
IP rating
IP20
Pollution degree rating
2 (with no aromatic compounds)
Altitude
< 2000 m

Demo

 

Quote

 
KRSS Left Strip