PAR-H2

The PAR-H2 series use:

- The latest polymer electrolyte membrane (PEM) technology to produce pure hydrogen.
- A single column dryer with programmable automatic regeneration via an integrated clever calendar.

The exclusive, electronically-controlled gas/liquid separator, automatic checking for internal leaks whenever starting the units and constant control of operating parameters, guarantee maximum safety.

The touch screen LCD interface provides simple and user-friendly management of all functions on the unit


  • Advantages
  • Characteristics
  • Specifications
  • Demo
  • Quote

Advantages

Improved chromatograph result
The use of hydrogen as a carrier gas allows lower temperature elution, thus extending the life of the chromatograph column. Hydrogen as a carrier gas is faster and more sensitive than the more-expensive helium. Run time savings of 25% to 35% without a decline in resolution.

Increased laboratory efficiency
A constant, uninterrupted gas supply of guaranteed purity eliminates interruptions of analyses to change cylinders and reduces the amount of instrument re-calibrations required.

Improved safety
The very limited internal volume (less than 50 ml) allows safe use of the gas generators where the use of cylinders is risky or prohibited.
The application of tested safety technologies stops the unit in the event of leaks or malfunctions.

Simple installation
Gas generators can be installed in the laboratory, on or under a bench, eliminating the need for long gas lines from cylinders secured elsewhere.

Characteristics

  • Patented gas/water separator electronically controlled
  • LCD touch screen : with indication in real time : H2 outlet pressure, H2 flowrate, water quality, water level, cartridge saturation, system status with auto-diagnostics of breakdown with alarms
  • Remote PC monitoring in standard via RS232 or USB to interface the unit with customers PC software (allow the end user to be connected with the manufacturer for remote diagnostics).
  • Water tank protected and filtered
  • Automatic checking for internal leaks to guarantee maximum safety
  • Automatic internal water tank refill included in standard
  • Capabilities allowing to work in parallel mode

Specifications

Models

PAR-H2-120

PAR-H2-180

PAR-H2-260

PAR-H2-400

PAR-H2-500

PAR-H2-650

General Information
H2 flow rate cc/min

120

180

260

400

500

650

H2 purity

> 99.9999 % (O2 < 0.1 ppm, dewpoint H2O < -70°C (-94°F))

Delivery pressure

1 - 12 bar (14 - 174 psi), 16 bar (232 psi) in option

H2 dryer

Single column dryer with programmable automatic regeneration via an integrated clever calendar

Internal water tank

2.3 liters with autorefill included (external water tank in option)

Temperature range

From 5 - 35°C (41° - 95°F) and humidity 80% to 25°C (77°F)

LCD touch screen

Resolution 128 x 64" touch screen with (operating parameters, system status, alarms)

Water quality

Deionised, ASTM II, < 0.1μ

Dimensions (W x H x D)

30 x 43 x 43 cm(12" x 17" x 17")

Outlet port

1/8 Swagelock

Weight (kg/lbs)

15.5 / 34

16.5 / 36

Power consumption

280 W

450 W

Power supply

90 - 240 V ac / 50 - 60 Hz

Communication
RS232

X

RS485/USB

Option

X

Special functions
Parallel mode

Option

X

Automatic internal water tank refill

X

Canister filling

X

Demo

 

Quote

 